Skip to main content

Automation at the service of standardization

C-CAD™ CONCEPT

Automatic process performed under horizontal agitation of the Platelet Concentrate bags enabling automatic transfer bag to bag fluid in closed circuit and temperature controlled.

C-CAD MAKES IT POSSIBLE

  • Ensure the step of residual amotosalen removing according the preparation of viro-attenuated platelet concentrates
  • Viro-attenuated platelet (DS ou LV)
  • Transfer of platelet products to a storage bag after a programmed period of CAD time
  • Agitation, clamp the transfer line before and after flow
  • Simultaneous treatment of 12 CP
  • Record the traceability of the elements and steps of the process

Benefits

High Flexibility in Production Planning

  • C-CAD creates more freedom to re-organize working conditions, allowing operators free time to tend to other tasks

Flexibility

  • 2 modes, “decay” or “CP retention,” in a single controller

Easy implementation and time saving

  • Many similarities with current platelet production processes make implementation of the C-CAD™ concept easy to perform

Standardization

  • The fully automated process assures a consistent quality

Efficiency

  • One C-CAD can process 12 PC simultaneously

Traceability

  • Creates a detailed record including process data with unit identifier for more complete process traceability

Compliance with European Directives

  • 2014/30/CE and EN 61326-1 2011 CEM class B

  • 2014/35/UE EN 61010-1:2011 and EN 61010-2-051:2015

  • 2011/65/CE RoHS

Specifications

  • Size (WxHxD): 871mm x 434mm x 396mm
  • Weight: 60kg
  • Voltage input: 220/240 VAC
  • Voltage Frequency: 50/60 Hz
  • Power: < 140W
  • Daily Consumption: 1.3kWh/24h
  • Heat emission: 9,5Kcal/h
  • Primary fuse: 250V T2A
  • Sound level: 40dbA à 1 mètre
  • Environment of use: Operating temperature: 15°C à 35°C
  • Working temperature: 0°C – 35 °C
  • Maximum altitude: 2000 m
  • Relative humidity < < 90% (non-condensing) 90% sans condensation
  • Storage conditions: -10°C à 50°C
  • Relative humidity < < 90% (non-condensing) 90% sans condensation
  • Pollution degree: 2
  • Protection degree: IPX0
  • Communication port: RS-232 for CAN connection
  • USB for software updates
    RJ-11: Connection of the agitation alarm to a CTM system

Accessories
AV power cord – Operating instructions

Optional:
PC (Professional Office Pack) – C-CAD softwares – Wireless barcode reader (support, cable).